STUDIES IN HYDRIDE GENERATION FOR THE DETERMINATION OF TRACE LEVELS OF LEAD BY ATOMIC FLUORESCENCE SPECTROMETRY | |
LUZ OLIVIA LEAL QUEZADA | |
Acceso Abierto | |
Sin Derechos Reservados | |
Spectrometry | |
Two methods were used for the determination of lead at ng L-1 levels by Flow Injection- Hydride Generation-Atomic Fluorescence Spectrometry (FI-HG-AFS). Lead hydride was generated using hydrogen peroxide and potassium hexacyanoferrate (III), both as oxidants agents. A reducing KBH4 solution, HCl solution and the sample were dispensed into a gas-liquid-separator. An argon flow delivered the lead hydride (PbH4) and the hydrogen gas into the flame of an atomic fluorescence spectrometer. Instrument parameters and chemical variables were optimized. A study of interference of the most common ions was also carried out. Analytical curves were obtained for lead determination with hydrogen peroxide between 1 and 80 g L−1, whereas using hexacyanoferrate (III) between 0.5 and 80 g L−1. The detection limit of the proposed technique was 0.42 g L−1 y 0.03 g L−1, respectively. The better results were observed with the hexacyanoferrate (III) as oxidant agent due its accuracy and sensitivity. | |