Por favor, use este identificador para citar o enlazar este ítem: http://cimav.repositorioinstitucional.mx/jspui/handle/1004/1433
STUDIES IN HYDRIDE GENERATION FOR THE DETERMINATION OF TRACE LEVELS OF LEAD BY ATOMIC FLUORESCENCE SPECTROMETRY
LUZ OLIVIA LEAL QUEZADA
Acceso Abierto
Sin Derechos Reservados
Spectrometry
Two methods were used for the determination of lead at ng L-1 levels by Flow Injection- Hydride Generation-Atomic Fluorescence Spectrometry (FI-HG-AFS). Lead hydride was generated using hydrogen peroxide and potassium hexacyanoferrate (III), both as oxidants agents. A reducing KBH4 solution, HCl solution and the sample were dispensed into a gas-liquid-separator. An argon flow delivered the lead hydride (PbH4) and the hydrogen gas into the flame of an atomic fluorescence spectrometer. Instrument parameters and chemical variables were optimized. A study of interference of the most common ions was also carried out. Analytical curves were obtained for lead determination with hydrogen peroxide between 1 and 80 g L−1, whereas using hexacyanoferrate (III) between 0.5 and 80 g L−1. The detection limit of the proposed technique was 0.42 g L−1 y 0.03 g L−1, respectively. The better results were observed with the hexacyanoferrate (III) as oxidant agent due its accuracy and sensitivity.
2010
Libro
Inglés
OTRAS
Versión revisada
submittedVersion - Versión revisada
Aparece en las colecciones: Libros

Cargar archivos:


Fichero Tamaño Formato  
BeltranCIPITECH2010.pdf399.12 kBAdobe PDFVisualizar/Abrir